GET A QUOTE

    OVERVIEW

    What you need to know

    X-ray Generator

    Complete solid-state high-frequency, high-voltage generator.

    Generator Type: High Frequency Inverter; 80 kHz

    Input PowerSingle phase 220VAC, 50/60Hz

    Radiographic Ratings: Large Focal Point 20-35kV/10-510mAs

    Small Focal Point 20-35kV/10-100mAs

    Power Rating6.2kW

    X-ray Tube:

    • Focal Spot SizeDual Focus 0.2 / 0.4mm
    • Target Material: Molybdenum (Mo)
    • Port Material: Beryllium (Be)
    • High-speed anode drive: 2800 /1000rpm
    • Target angle:12°/12°
    • Anode Heat Storage100KJ150KHU
    • Anode Cooling: Air cooling
    • FiltrationMo0.03mm),Al0.5mm)

    Radiographic Stand:

    18×24cm (24cmX30cm optional)

    Bucky Device18×24cm Bucky drive mechanism,(24cmX30cm optional)

    Grid Ratio5:1 30 Line/cm

    Radiographic Stand:

    18×24cm (24cmX30cm optional)

    Bucky Device18×24cm Bucky drive mechanism,(24cmX30cm optional)

    Grid Ratio5:1 30 Line/cm

    Line Voltage:

    220Vac@25A,Single phase

    Weight:

    240kgGross weight 340kg

    W×H: 940mm×560mm×2000mm